型號 | 多通道耐壓測試儀 19053 | |
模式 | ACV/DCV/IR/DCR/SCAN/IWT(optional) | |
耐壓測試 | ||
輸出電壓 | AC:0.05~5KV,DC:0.05~6KV | |
加載分辨率 | 1% of setting+0.1% of full scale | |
電壓分辨率 | 2V | |
電壓準(zhǔn)確度 | 1% of setting+0.1% of full scale | |
切斷電流 | AC:30mA,DC:10mA | |
電流分辨率 | AC:1μA,DC:0.1μA | |
電流準(zhǔn)確度 | 1% of reading+0.5% of range.(1% of reading+5% of total current) | |
輸出頻率 | 50Hz/60Hz | |
測試時(shí)間 | 0.3~999sec,continue | |
上升時(shí)間 | 0.1~999 sec,off | |
下降時(shí)間 | 0.1~999 sec,off | |
停滯時(shí)間 | 0.1~999 sec,off | |
信號模式 | 單信號 | |
絕緣電阻測試 | ||
輸出電壓 | DC:0.05~5KV | |
電壓解析度 | 2V | |
電壓準(zhǔn)確度 | 1% of setting+0.1% of full scale | |
IR范圍 | 1MΩ~50GΩ | |
電阻解析度 | 0.1MΩ | |
阻抗準(zhǔn)確度 | <500V | 0.1MΩ~1GΩ:±3% of reading+(0.2*500/Vs)% of full range |
≥500V <1000V | 1MΩ~1GΩ:±3% of reading+0.1% of full range 1GΩ~10GΩ:±7% of reading+2% of full range 10GΩ~50GΩ:±10% of reading+1% of full range | |
>1000V | 1MΩ~1GΩ:±3% of reading+0.1% of full range 1GΩ~10GΩ:±7% of reading+2% of full range 10GΩ~50GΩ:±10% of reading+1% of full range | |
掃描器 | Ports,±phase |
| |||||||||||||||||||||||||||||
|
Test Voltage | 0 - 3000 VAC True RMS 100 - 4250 VDC | Accuracy | ± 1% Full Scale |
Input Current | 1 A | Weight | 15 lbs |
Input Frequency | 50 - 60 Hz | Input Voltage | 100 - 200 or 220 - 240 VAC (Specify when ordering) |
Test Time | 1 sec or 60 sec (selectable)++ | Dimensions | 12" x 11" x 5" |
Failure Indication | ARC detection / Trip current limit 10 mA |
型號 | 9032C | 耐壓測試儀19032 |
模式 | DCV/ACV/IR/GB/LC(option) | |
耐壓測試 | ||
輸出電壓 | AC:0.05~5kV,DC:0.05~6kV | |
加載分辨率 | 1%+5V | |
電壓分辨率 | 2V | |
電壓度 | 1%+5counts | |
切斷電流 | AC:40mA,DC:12mA | |
電流分辨率 | AC:1μA,DC 0.1μA | |
電流度 | 1%+5counts | |
頻率 | 50Hz/60Hz | |
測試時(shí)間 | 0.3~999sec,continiu | |
上升時(shí)間 | 0.3~99.9sec,off | |
信號模式 | 單信號 | |
絕緣測試 | ||
輸出電壓 | DC:0.05~1kV | |
電壓分辨率 | 2V | |
電壓度 | 2%+5V | |
IR范圍 | 1MΩ~50G | |
阻抗分辨率 | 0.1MΩ | |
阻抗度 | 5% | |
接地電阻測試 | ||
輸出電流 | AC:3~30A | |
電流分辨率 | 10mA | |
電流度 | 1%+5counts | |
GR范圍 | 10~510mΩ | |
阻抗分辨率 | 0.2mΩ | |
阻抗度 | 1%+5counts | |
測試模式 | 4線測量 | |
補(bǔ)償功能 | 0~100mΩ | |
ARC探測 | ||
設(shè)置模式&, amp;, lt;, SPAN lang=EN-US style="FONT-SIZE: 9pt"> | 可程設(shè)置 | |
探測電流 | AC:1~40mA,DC:1~12mA | AC:1~20mA,DC:1~12mA |
最小脈寬 | 10μs approx | 40/20/10/4μs selective |
安全保護(hù)功能 | ||
快速輸出切斷 | 0.4ms(NG發(fā)生后) | |
快速直流釋放 | <0.2sec | |
鎖定 | 密碼鎖定 | |
GO/NG判定 | ||
報(bào)警 | GO:短鳴;NG:HI,LOW,ARC NG 長鳴 | |
Data Hold | 最近測試數(shù)據(jù)記憶 | |
Step Hold | 單步觸發(fā) ON/OFF | |
記憶存儲 | 10步50組呼叫 | 50步100組呼叫 |
自動泄漏電流掃描器(可選) | ||
模式 | LC/VA | |
泄漏電流 | ||
DUT輸入電源 | AC:300V/10A(最大) | |
短路保護(hù) | 10A/250V(保險(xiǎn)絲) | |
電流范圍 | 0.1~590μA/500~6000μA | |
電流分辨率 | 0.2μA/2μA | |
電流度 | 2%+5counts(DC-5kHz) | |
測量模式 | ||
測量模式 | Normal,Reverse,Single fault normal,Single fault reverse | |
測量設(shè)備 | UL554NP:Medical appliances& equipment UL554P:Medical appliances& equipment UL 1563:Electrical equipment & component UL 2601,IEC 601-1,EN6060-1:Medical appliances & equipment UL3101,UL1950,UL950,,IEC1010,IEC950,EN60950:Experiment & information equipment 5 different kinds human body RC network | |
電壓表 | ||
電壓范圍 | AC:0~300V | |
電壓度 | 1.5%+5counts | |
電流表 | ||
電流范圍 | AC:0.1~10A | |
電流度 | 1.5%+5counts | |
上下限 | ||
LC上下限 | 0~9.99mA,1μA(解析度) | |
電流上下限 | 0~9.999A | |
VA上下限 | 0~2200V | |
VA分辨率 | 0.1VA | |
一般規(guī)格 | ||
測試環(huán)境 | 溫度:0℃~40℃ 濕度:≤80%RH | |
功率 | 空載:100W,額定負(fù)載:>650W | |
電源 | 100V/120V/220V(AC±10%)/240V(AC+5%~-10%),50/60Hz | |
重量 | 約24 kg | |
尺寸(W×H×D) | 430×133×470mm |
交直流耐壓測試儀 型號:CJ24JK2673A | 貨號: |
交直流耐壓測試儀特點(diǎn): l 手動、自動測試 2 測試時(shí)間任意設(shè)定 3 切斷電流任意連續(xù)設(shè)定 4 擊穿保護(hù)響應(yīng)快 5 不合格聲、光報(bào)警 交直流耐壓測試儀主要技術(shù)指標(biāo): 型號 參數(shù) CJ24JK2673A 輸出電壓 AC/DC 0—5KV 精 度 電壓:±3% 電流:±5% 切斷電流 AC: 0—2/20mA DC: 0—2/20mA 測試時(shí)間 0—99秒或手控 顯示方式 電壓、電流、時(shí)間全數(shù)顯 體 積 315mm×260mm×168mm | 產(chǎn)品圖片 |
PTS系列攜帶式直流測試器組包含了直流耐高壓測試器及高阻計(jì)雙功能,只要一部儀器即可測試介電強(qiáng)度及絕緣電阻;專門設(shè)計(jì)用于現(xiàn)場操作,具有電氣維護(hù)測試之各種優(yōu)點(diǎn)。 ◆ 提供37.5KV,75KV,80 KV,100KV,130KV,200KV 6種機(jī)型,額定電流最大10mA ◆ 5個檔位之電流表,具有0-1μA,解析度10nA,用于泄漏電流量測 ◆ 在任何輸出電壓下,高阻量測可達(dá)7,500GΩ ◆ 良好調(diào)整輸入能力,在±10%電源變動下為±1%穩(wěn)定度,適合現(xiàn)場使用 ◆ 可連續(xù)操作 ◆ 適用纜線絕緣、開關(guān)、馬達(dá)、變壓器、瓷瓶、發(fā)電機(jī)及其它設(shè)備 ◆ 安全保護(hù)功能:泄漏電流旁路選擇、突波保護(hù)、外部安全互鎖、零電位啟動 | ||||||||||||||||||||||||||||||||||||||||||
技術(shù)參數(shù) | ||||||||||||||||||||||||||||||||||||||||||
共同規(guī)格 指示表檔位: 0~1.0μA電流,倍數(shù)×1.0,×10,×100,×1k,×10k 1~100MΩ電阻,倍數(shù)×0.1,×1,×10,×100,×1K 指示表型式:3.5寸指針式,2%精度 標(biāo)準(zhǔn)附件 (1)20尺RG8/U同軸輸出導(dǎo)線PTS130/200與X型 (2)20尺回路線 (3)20尺接地線 (4)10尺電源線 (5)20尺短路接地線
|